SLAUGHTER 2630 CALIBRATION and SLAUGHTER 2630 REPAIR

 
A calibration by Custom-Cal is performed by engineers with extensive OEM experience. We have the expertise and the necessary standards to perform the SLAUGHTER 2630 Calibration, onsite calibration may be available. We specialize in quick turnaround times and we can handle expedited deliveries upon request.

 

 
   SLAUGHTER 2630   Description / Specification:    
SLAUGHTER 2630 30A AC Ground Bond Tester

The Slaughter 2630 is an enhanced Ground Bond tester that is a companion instrument to the 2900 series. Whether used with the 2900 series as a system or as a stand alone instrument, the 2630 is equipped with the technology and features which make it a perfect choice for production line safety testing. It features Milliohm Offset Capability that minimizes the effect of test lead resistance or resistance of test fixturing. The milliohm offset function allows the resistance of longer test leads and test fixtures to be subtracted from the test results.

Specifications

Output
Current: AC 3 - 30 Amps, 0.1 Amp/step
Regulation: ± (2% of Setting + 0.02A)
Voltage: AC 6 Volts fixed

Output Frequency: 50/60 Hz user selectable

Dwell Time Setting
0 and 0.5 - 999.9 seconds, 0.1 second/step
0 for continuous running
Accuracy: ± (0.1% of Setting + 0.05 seconds)

Failure Settings
High limit: 0 - 120mΩ for 10 - 30 Amps, 1mΩ/step
0 - 510mÙ for 10 Amps, 1mÙ/step
Accuracy: ± (2% of setting + 2mΩ)

Milliohm Offset
Capability: 100mΩ, 1mΩ/step
Accuracy: ± (2% of setting + 2mΩ)

Metering
Ammeter (3 digits)
Range: 0 - 30 Amps
Resolution: 0.1 Amp/step
Accuracy: ± (3% of Reading + 0.03A)
Ohmmeter (3 digits)
Range: 0 - 510mΩ
Resolution: 1 mΩ/step
Accuracy: + (2% of Reading +2mΩ)
Timer (4 digits)
Range: 0 - 999.9 seconds
Resolution: 0.1 seconds/step
Accuracy: ± (0.1% of Reading + 0.05 seconds)



 

Standard Calibration $635.00 *
*This is a Web introductory price for one calibration of the SLAUGHTER 2630. Price does not in most cases include measurement performance data. Pricing does include NIST traceable calibration and issue of a calibration certificate and calibration label. Pricing may vary slightly due to volume and location of laboratory supporting calibration. Volume pricing may apply. On-site fees may apply depending on logistics, location and volume of work to be completed during the visit.


Related Bench Equipment Terms and Definitions. For a complete list go to our  Terms and Definitions Page.

Available Gain
Available Gain is the ratio, at a specific frequency, of power available from the output to the power in from the source.

Decibel (dB)
A unit of measurement of power which indicates relative power. A -10 dB means a reduction in power by 10 times, -20 dB means another 10 times or 100 times overall, -30 means another 10 times or 1000 times overall and so on.

Mismatch Uncertainty
Mismatch Uncertainty is caused by re-reflections between one device (the source) and the device that follows it (the load).

Pulse Power
Pulse Power, the energy transfer rate is averaged over the pulse width. Pulse width is considered to be the time between the 50 percent risetime/falltime amplitude points. Pulse power averages out any aberrations in the pulse envelope such as overshoot or ringing.


Please contact us for your SLAUGHTER 2630 CALIBRATION and/or SLAUGHTER 2630 30A AC Ground Bond Tester REPAIR

I'm interested in: 
Calibration Type (Select one):
Repair Information:
 Serial Number:

 Is there any physical damage?  
 If yes, describe:


What is the description of the problem?

How/when did the problem start?


 (*) Required Fields    For formal quote please fill in all fields.
Company:
From (Email):*
Contact Name:*
Country:*
Address 1:
Address 2:
City:
State/Territory/Province: *
Zip/Postal Code:
Phone:*

Comment: