Keysight (Agilent) B1517A REPAIR and Keysight (Agilent) B1517A CALIBRATION

 
Custom-Cal has a high success rate in the repair of the Keysight (Agilent) B1517A. A calibration by Custom-Cal is performed by engineers with extensive OEM experience. We have the expertise and the necessary standards to perform the Keysight (Agilent) B1517A Calibration, onsite calibration may be available. We specialize in quick turnaround times and we can handle expedited deliveries upon request.

Need another Semiconductor Tester?
Shop for Comparable Products on our
 used Semiconductor Tester page

 
   Keysight (Agilent) B1517A   Description / Specification:    
Keysight (Agilent) B1517A  High resolution source/monitor unit

The Agilent B1517A High resolution source/monitor plug-in module features a range up to 100 V/0.1 A with 4-quadrant operation. Minimum measurement resolution 1 fA/0.5 μV. Optional ASU (atto-sense and switch unit) for 100 aA resolution and IV/CV switching capability. It uses one slot in a B1500A Mainframe.
Specifications.

Voltage (high resolution ADC).
Range, Force resolution, Measure resolution, Force accuracy ±(% read value + offset voltage V), Measure accuracy (% read value + offset voltage V).
  ±0.5 V, 25 μV, 0.5 μV, ±(0.018% + 150 μV), ±(0.01% + 120 μV).
  ±2 V, 100 μV, 2 μV, ±(0.018% + 400 μV), ±(0.01% + 140 μV).
  ±5 V, 250 μV, 5 μV, ±(0.018% + 750 μV), ±(0.009% + 250 μV).
  ±20 V, 1 mV, 20 μV, ±(0.018% + 3 mV), ±(0.009% + 900 μV).
  ±40 V, 2 mV, 40 μV, ±(0.018% + 6 mV), ±(0.01% + 1 mV).
  ±100 V, 5 mV, 100 μV, ±(0.018% + 15 mV), ±(0.012% + 2.5 mV).

Current (high resolution ADC).
Range, Force resolution, Measure resolution, Force accuracy ±(% + A + A), Measure accuracy (% + A + A), Max voltage.
  ±10 pA, 5 fA, 1 fA, ±(0.46%+30 fA+10 aA x Vo), ±(0.46%+15 fA+10 aA x Vo), 100 V.
  ±100 pA, 5 fA, 2 fA, ±(0.3%+100 fA+100 aA x Vo), ±(0.3%+30 fA+100 aA x Vo), 100 V.

Power consumption.
Voltage source mode (Ic is the current compliance setting):
  20 x Ic (W) (0.5 to 20V),
  40 x Ic (W) (40V),
  100 x Ic (W) (100V).
Current source mode (Vc is the voltage compliance setting and Io is output current):
  20 x Io (W) (Vc ≤ 20),
  40 x Io (W) (20 < Vc ≤ 40),
  100 x Io (W) (40 < Vc ≤ 100).



 



Related Bench Equipment Terms and Definitions. For a complete list go to our  Terms and Definitions Page.

Carrier Wave (CW)
Carrier Wave or Continuous Wave is a sinusoidal signal without modulation.

Harmonic
A harmonic of a wave is a component frequency of the signal that is an integer multiple of the fundamental frequency, i.e. if the fundamental frequency is f, the harmonics have frequencies 2f, 3f, 4f, . . . etc.

Noise Floor
Noise floor is the measure of the signal created from the sum of all the noise sources and unwanted signals within a measurement system.

Ripple
Ripple is the small unwanted residual periodic variation of the direct current (dc) output


Please contact us for your Keysight (Agilent) B1517A REPAIR and/or Keysight (Agilent) B1517A High resolution source/monitor unit CALIBRATION

I'm interested in: 
Calibration Type (Select one):


Repair Information:
 Serial Number:

 Is there any physical damage?  
 If yes, describe:


What is the description of the problem?

How/when did the problem start?


 (*) Required Fields    For formal quote please fill in all fields.
Company:
From (Email):*
Contact Name:*
Country:*
Address 1:
Address 2:
City:
State/Territory/Province: *
Zip/Postal Code:
Phone:*

Comment: