The Agilent 85032E type-N calibration kit is used to calibrate Agilent network analyzers up to 6 GHz for measurements of components with 50 ohm type-N connectors. The 85032E calibration kit contains the following: one male combination open/short termination and one male 50W load. The broadband loads are instrument-grade, 50W terminations that have been optimized for performance up to 6 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss. The opens and shorts are built from parts that are machined to the current state-of the-art precision machining. The short’s inner conductors have a one-piece construction, common with the shorting plane. This construction provides for extremely repeatable connections. Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. Some of the opens and shorts have offsets. The lengths of these offsets are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.
|