FLUKE 5500A-COIL REPLACEMENT / ALTERNATE

Custom-Cal has a solution to replace your FLUKE 5500A-COIL. We specialize in quick turnaround times and we can handle expedited deliveries upon request.

Alternative current coil.
Original
Replacement
  List Price $1,372.00
FLUKE 5500A-COIL 50 Turn Current Coil

TRANSMILLE EA002 Clamp Coil Adapter

FLUKE 5500A-COIL   Description / Specification: TRANSMILLE EA002   Description / Specification:      
The Fluke 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrating clamp-type current meters that operate by two different principles-as current transformers (AC only), and by the Hall Effect (Both AC and DC).
Specifications.

Number of Turns: 50.
Maximum Current: 11A rms, continuous (20.5A rms, 2 minutes).
Maximum Duty Cycle Derating: < 11A, continuous (> 11A, 2 minutes ON, 8 minutes OFF).
Maximum Voltage: 7 V dc or 7 V rms ac.
Minimum Inner Diameter of Clamp Jaws: 2.54 cm [1”].


The Transmille EA002 Clamp Coil Adapter is designed for the calibration of both wound (AC only) & Magnetic field (hall effect) AC/DC clamp meters the current coil offers several unique features built in a robust construction. The coil is designed to immerse both jaws of the clamp meter in the magnetic field while allowing the ‘gap’ or opening where magnetic flux will escape to protrude through the coil and stay out of the strongest part of the field. This makes the reading less dependent on the position of the clamp within the coil and also the quality of the jaw closing, allowing greater confidence in the calibration.
Specifications.

Connection to Calibrator: Via supplied 4mm to 4mm connection leads (x2) to 3000 series current terminals.
Coil Configuration: 2 Turn (LHS), 10 Turn (RHS), 50 Turn (CENTRE).
Coil Type: High accuracy balanced configuration.
Min. internal jaw dimensions: 10mm (2 Turn & 10 Turn), 25mm (50 Turn).
Maximum Current: 30A.
Maximum RMS voltage: 4V.
Frequency Range: DC to 500Hz.

2 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 60A).
Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.35% + 0.008A.
Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.48% + 0.07A.

10 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 300A).
Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.41% + 0.01A.
Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.59% + 0.11A.

50 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 1500A).
Effective accuracy - Coil only (wound clamps, 90 Day Rel.) 0.24% + 0.04A.
Effective accuracy - Coil only (hall effect clamps, 90 Day Rel.) 0.45% + 0.42A.

Please contact us about the TRANSMILLE EA002 as a replacement for the FLUKE 5500A-COIL.